Scanning in Hand-Held Mode

1. Connect the scanner cable to the ACT Plus
beep when power is on.
2. Aim the scanner at the bar code and press the trigger.
3. Be sure that the red line from the scanner goes across the entire bar code.
When the scanner has successfully decoded the bar code, the scanner will beep,
and the information will be entered into the ACT Plus

Setting Bar Code Format Options

The scanner is preset to read UIDs and PIDs in format Code 39 With Check Digit
(Mod 10), Interleave 2 of 5 with Check Digit (Mod 49), Codabar and Code 128. It is
also preset to read the cartridge/ control information.
To set additional barcode format options:

Code 39

Scan the "Enable Code 39 Check Digit" barcode if the UID/PID barcodes use Code
39 check digit to verify a correct scan.
Scan the "Disable Code 39 Check Digit" barcode if the UID/PID barcodes do not
use Code 39 check digit to verify a correct scan. The default setting is Code 39
Disabled, so this barcode should only be used if Code 39 was mistakenly enabled
in the previous step.
Note: Disabling the Code 39 Check Digit will allow the scanner to scan a barcode
that has the check digit and add it to the data.
62 English
Figure 9. Correct scan location
Figure 10. Enable Code 39 Check Digit
Figure 11. Disable Code 39 Check Digit
®
instrument. The scanner will
®
instrument.
®