When there are 2 lot numbers entered for a cartridge type, the following variable
function keys will appear: [Toggle Active], [Remove Lot], [Edit Lot/Exp Date] and
[Exit to Main Menu].
To change the lot number of the currently active cartridge:
1. Select [Toggle Active].
*
2. Move the "
" to the cartridge lot number that is currently in use.
To remove a cartridge lot:
1. Select [Remove Lot].
2. To select the lot to be removed, use the up/down arrows (↑↓) to toggle to the
appropriate lot number.
3. To delete the selected cartridge lot number, select [Remove Selected Lot].

Bar Code Entry

From the Main Menu:
1. Select [Cartridge Lot]. Page 1 of the Cartridge Lot/Expiration Date screen
will appear.
When there is no or 1 lot number for a cartridge type:
1. Scan the bar code on the cartridge box. The lot number and expiration date
will automatically populate their respective fields.
2. To return to the Main Menu, press [Main Menu].
When 2 lot numbers exist for a cartridge type:
1. Select [Remove Lot].
2. To remove 1 of the lot numbers, toggle to the appropriate lot number.
3. To delete the selected cartridge lot number, select [Remove Selected Lot].
4. Scan the bar code on the cartridge box. The lot number and expiration date
will automatically populate their respective fields.
5. To return to the Main Menu, select [Main Menu].

View Test Results for the Last 20 Tests Performed

1. To display the results for the last 20 tests, select [View Test Results].
2. Use the up/down arrows (↑↓) on the left and right of the screen to scroll
through the test data.
3. The date, time, patient ID and test results will be displayed.
Note: If the ACT Plus
however, they are not erased from the result log.
28 English
®
power is cycled, the ability to display previous results is lost;