Error Messages - Medtronic ACT Plus Manuel De L'utilisateur

Masquer les pouces Voir aussi pour ACT Plus:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 81

Error Messages

Disk error - Disk is faulty.
Serial port error.
Motor stall. Power OFF then ON (E9).
System error. Power OFF then ON.
Bad Reference in Ch1 and Ch2.
OR
Bad Reference in Ch1.
OR
Bad Reference in Ch2.
Error: E1
Error: E2 or E3
No flags.
OR
No flag in Ch1.
OR
No flag in Ch2.
OR
No flags in Ch1 or Ch2.
Insufficient data.
Error: Invalid reply.
Light path failure in Ch1 and Ch2 (E4).
OR
Light path failure in Ch1 (E4).
OR
Light path failure in Ch2 (E4).
Motor stall.
Motor stall, cycle power (E8).
No reply.
File system error.
Comm. Error; Invalid data.
®
ACT Plus
Operator's Manual
®
The ACT Plus
instrument is trying to read a
floppy disk or USB storage device and
cannot read it. Insert a floppy disk or USB
storage device that meets the specification
listed in Section 11, Specifications.
An error occurred in accessing the serial
port. Retry transfer.
The motor has been inhibited. To reset the
system, turn the power OFF then ON.
The system has become unstable. To reset
the system, turn the power OFF then ON.
®
The ACT Plus
instrument could not get a
baseline drop time for Channel 1, Channel 2,
and both channels. Repeat test.
No cartridge is detected. Place a cartridge in
the actuator assembly and retest.
A software error has occurred. Call
Medtronic Instrument Service.
®
The ACT Plus
instrument did not sense a
flag in Channel 1, Channel 2, or both
channels. Repeat the test.
The channels of the control timed out at 999.
Repeat the control.
Internal communication error. Call Medtronic
Instrument Service.
The flag sensor has failed. Clean the
actuator assembly cover per the instructions
in the Maintenance section of this manual. If
the error is not corrected, call Medtronic
Instrument Service.
The motor is inhibited. Press STOP to reset.
Repeat the test.
The motor is inhibited. To reset, power OFF
then ON.
Internal communication error. Call Medtronic
Instrument Service.
Call Medtronic Instrument Service.
Call Medtronic Instrument Service.
English 53

Publicité

Table des Matières
loading

Table des Matières