Set the QC Interval for Liquid and Electronic Controls:

From the QC Manager Setup screen:
1. Select [QC Interval Setup]. The following variable function keys appear:
[Liquid Interval], [ACTtrac

To set the liquid QC interval:

1. Select [Liquid Interval].
2. To view the choices, [None], [8 hours] or [7 days], continue to press [Liquid
Interval].
3. To confirm the choice, press [Enter].
To set the ACTtrac
1. Select [ACTtrac
2. To view the choices, [None] or [8 hours], continue to press [ACTtrac
Interval].
3. To confirm the choice, press [Enter].

Enable or Disable Test Cartridge Types:

From page 2 of the QC Manager Setup screens:
1. Select [Cartridge Enable Setup].
2. To select the desired cartridge, use the arrows on either side of the display of
cartridges to navigate up or down.
3. Select the desired cartridge.
4. To view the choices, [ON] or [OFF], continue to press the desired cartridge.
5. To confirm the choice, press [Enter].
Download All of the Instrument Settings to a Floppy Disk or USB
Storage Device:
From page 2 of the QC Manager Setup screens:
1. Insert a formatted disk or USB Storage Device into the appropriate drive.
2. Select [Download Settings].
Set the Permanent Record or the Mode by Which the ACT Plus
Instrument Maintains a Record of Downloaded Data:
From page 2 of the QC Manager Setup screens:
1. Select [Perm Record].
2. To view the choices, [None], [Floppy], [USB] or [Electronic], continue to
press [Perm Record].
38 English
®
Interval] and [Exit to QC Manager Setup].
®
electronic QC interval:
®
Interval].
ACT Plus® Operator's Manual
®
®