Notes: If a System Alarm occurs that cannot be resolved by cycling the power OFF
then ON, record the 3-digit Error Code (if displayed), and contact Medtronic
Instrument Service for assistance.
Temperature
Heat Block Temp XX.X°C.
Heat Block Temp > 44°C (E99).
Heat Block Temp < 5°C (E00).
54 English
The current actuator heat block temperature
(XX.X°C) is displayed when it is between
20°C and 41°C.
The actuator heat block temperature is
greater than 44°C.
The actuator heat block temperature is less
than 5°C.
ACT Plus® Operator's Manual