CALYS 150
C.5 Calibration procedure
The CALYS 150 is capabable of creating a calibration report from a pre-defined procedure.
The number of procedures that can be recorded depends on the size of the available memory and the size of each procedure
(number of test points). If the memory is not being used by other functions, it is possible to record several tens of procedures.
To find the available memory size, refer to the chapter, Storage of current acquisitions (chapter C.6).
Open the MEASUREMENT MENU dialog box,
Select the Calibration procedure function,
Press ENTER.
Use the Up
/ Down
new procedure. A procedure may be duplicated with the F2 key (Duplicate). The F4 key is used to delete a selected
procedure.
Press ENTER.
With edit keys, enter "Device name" which is displayed in the "Reference" field in the above list.
Enter "Manufacturer" name which is displayed in the "Manufacturer" field in the above list.
The instrument features 2 calibration methods: "By comparison" or "Standard generator". Select from drop-down list.
In both methods, select "Device measurement": "Channel 1" or "Keyboard entry".
170 / 194
NTA47224-300A9
arrow buttons to adjust the parameters or press the function button F1 (new) to define a
ADVANCED OPERATION