During generation of a cyclic ramp, a progress bar indicates the state of progress. The function keys are used to control
generation:
The
key stops generation at any time
The
key suspends generation
The
key commences or resumes generation
An
icon in the transmission window indicates suspended generation.
Synthesizer Mode
The synthesizer function is used:
to store up to 100 transmission values in permanent memory,
to recall and transmit the contents of these memories manually or automatically.
When this mode is active the
The
key is used to launch generation of values in increasing order and the
generation of values in decreasing order.
The default parameters of this mode are displayed in the left hand side of the transmission window.
The parameters of the synthesizer mode are:
First point no.: number of the first point in a cycle
Last point no.: number of the last point in a cycle
T: the duration for which a point is transmitted.
Nbr: the number of polling cycles
Delay: delay between launch and transmission of the first point.
To change these parameters, use the Menu
The number of the first point may be higher than that of the last point. All points between the first and last are generated.
icon appears in the transmission window.
Synthesizer...
Simulation Modes
function key is used to launch
Parameters...
NTA47224-300A9
C