The settling time field can be used to define the time, in seconds, between sending the setting from the output of the CALYS
150 and making the measurement at its input.
The unit may be set to display a "Verdict": "OK" or "KO" at the end of the procedure. In this case, set permissible deviation
in percent and units (according to the type of measurement).
When all fields have been completed, press ENTER.
When editing is completed, exit Menu with ESC key.
To run a calibration procedure, return to "Calibration procedure" menu and select "Run".
Calibration procedures
NTA47224-300A9
C