CALYS 150
In the case of the 0-20 mA range, the scale curve is as follows:
In the case of the 4-20 mA range, the scale curve is as follows:
The CALYS 150 displays in the window details of the selected configuration using the following icons:
: to show loop power supply off
: to show loop power supply on
: to show square law scale
: to show HART compatibility.
When the values measured are near the lower limit of the range (0 mA or 4 mA) a small variation in the values
measured translates into a more significant variation for the converted values (in %) because of the square law
nature of the scaling.
124 / 194
NTA47224-300A9
20
18
16
14
12
10
8
6
4
2
0
0
5
10
x (mA)
20
18
16
14
12
10
8
6
4
2
0
0
5
10
x (mA)
GETTING STARTED
15
20
15
20