4 Using the Regenerative Power System
The menu to unlock the front panel appears every time a key is pressed. Enter the password to unlock
the front panel.
If the password is lost, the
front panel lockout password. Refer to
The
SYSTem:COMMunicate:RLSTate RWLock
command is completely independent of the front panel lockout function. If you use this command to
lock the front panel, the front panel will be unlocked when AC power is cycled.
Password Protection
You can password-protect all functions located in the Admin menu. These include: instrument
calibration, interface access, non-volatile memory reset, firmware update, password updates.
As shipped, the Admin menu password is 0 (zero). This means that you do not have to enter a password
to access the Admin menu. Simply select System\Admin\Login and press Enter. To password-protect
the Admin menu:
Front Panel Menu Reference
Select System\Admin\Password
Enter a numeric password, up to 15 digits long. Then
press Select.
Log out of the Admin menu to activate the password. You
can now only enter the Admin menu by providing the
correct password in the Password field.
If the password is lost, access can be restored by setting an internal switch to reset the password to 0.
If the message "Locked out by internal switch setting" or "Calibration is inhibited by switch setting"
appears, the internal switch is set to prevent the password from being changed. Refer to
Switches
for more information.
Clock Setup
The real-time clock is used to provide time stamp information for the Keysight 14585A Control and
Analysis Software, which is its only function. When shipped, the real-time clock is set to Greenwich
mean time. To set the clock:
Front Panel Menu Reference
Select System\Preferences\Display\Clock.
Enter the date in the Month, Day and Year fields.
Enter the time in the Hour, Minute, and Second fields.
Press Select to set the date and time.
174
SYSTem:PASSword:FPANel:RESet
Calibration Switches
command can also lock and unlock the front panel. This
SCPI Command
Enter calibration mode using the original password
CAL:STAT ON, <password>
To change the password:
CAL:PASS <password>
To exit calibration mode and activate the password:
CAL:STAT OFF
SCPI Command
To set the date:
SYSTem:DATE 2018,06,30
To set the time:
SYSTem:TIME 20,30,0
Keysight RP7900 Series Operating and Service Guide
command can reset the
for more information.
Calibration