6 Verification and Calibration
Keysight RP7952A/RP7962A
RP7952A/RP7962A
Test Record
Test Description
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
CV Load Effect:
CV Ripple and Noise
peak-to-peak:
rms:
Transient response @ 500 µs:
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
CC Load Effect:
Current Sink Tests
100% of current rating:
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
CC Load Effect:
100% of current rating (Isink):
292
Report Number ______
Model
Min. Specs
Both
0.43985
Both
Vout - 0.08015
Both
499.78
Both
Vout - 0.230
Both
- 0.030
Both
N/A
Both
N/A
Both
- 1.25
Both
- 0.024
Both
Iout - 0.024
Both
39.936
Both
Iout - 0.064
Both
- 0.017
Both
- 40.064
RP7952A/RP7962A UUT Settings
Voltage priority: 0.5 V, 20.5 A
Voltage priority: 500 V, 20.5 A
Voltage priority: 250 V, 41 A
Voltage priority: 250 V, 41 A
Current priority: 255 V, 0 A
Current priority: 255 V, 40 A
Current priority: 510 V, 20 A
Current priority: 255 V, - 40 A
Keysight RP7900 Series Operating and Service Guide
Date__________
Results
Max. Specs
_______
0.56015
_______
Vout + 0.08015
_______
500.21
_______
Vout + 0.230
_______
+ 0.030
_______
0.500
_______
0.100
_______
+ 1.25
_______
0.024
_______
Iout + 0.024
_______
40.064
_______
Iout + 0.064
_______
+ 0.017
_______
- 39.936
RP7952A/RP7962A Load Settings
not used
not used
Current priority: 255 V, -40 A
Current priority: 285 V, -20 A to -40 A, slew=max
Voltage priority: 250 V, 0 A
Voltage priority: 250 V, -41 A
Voltage priority: 500 V, -20.5 A
Voltage priority: 250 V, -41 A