Test Record Forms
Keysight RP7951A/RP7961A
RP7951A/RP7961A
Test Record
Test Description
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
CV Load Effect:
CV Ripple and Noise
peak-to-peak:
rms:
Transient response @ 500 µs:
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
CC Load Effect:
Current Sink Tests
100% of current rating:
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
CC Load Effect:
100% of current rating (Isink):
Keysight RP7900 Series Operating and Service Guide
Report Number ______
Model
Min. Specs
Both
0.43985
Both
Vout - 0.08015
Both
499.79
Both
Vout - 0.230
Both
- 0.030
Both
N/A
Both
N/A
Both
- 1.25
Both
- 0.012
Both
Iout - 0.012
Both
19.968
Both
Iout - 0.032
Both
- 0.009
Both
- 20.032
RP7951A/RP7961A UUT Settings
Voltage priority: 0.5 V, 10.25 A
Voltage priority: 500 V, 10.25 A
Voltage priority: 250 V, 20.5 A
Voltage priority: 250 V, 20.5 A
Current priority: 255 V, 0 A
Current priority: 255 V, 20 A
Current priority: 510 V, 10 A
Current priority: 255 V, - 20 A
6 Verification and Calibration
Date__________
Results
Max. Specs
_______
0.56015
_______
Vout + 0.08015
_______
500.21
_______
Vout + 0.230
_______
+ 0.030
_______
0.500
_______
0.100
_______
+ 1.25
_______
0.012
_______
Iout + 0.012
_______
20.032
_______
Iout + 0.032
_______
+ 0.009
_______
- 19.968
RP7952A/RP7962A Load Settings
not used
not used
Current priority: 255 V, -20 A
Current priority: 285 V, -10 A to -20 A, slew=max
Voltage priority: 250 V, 0 A
Voltage priority: 250 V, -20.5 A
Voltage priority: 500 V, -10.25 A
Voltage priority: 250 V, -20.5 A
291