SENSe Subsystem
Sense commands control the current measurement ranges and window as well as the data acquisition
sequence.
SENSe:AHOur:RESet
SENSe:WHOur:RESet
Resets the amp-hour or watt-hour measurement to zero.
Parameter
(none)
Resets the amp-hour measurement: SENS:AHO:RES
Resets the watt-hour measurement: SENS:WHO:RES
SENSe:ELOG:FUNCtion:CURRent 0|OFF|1|ON
SENSe:ELOG:FUNCtion:CURRent?
SENSe:ELOG:FUNCtion:VOLTage 0|OFF|1|ON
SENSe:ELOG:FUNCtion:VOLTage?
Enables or disables the Elog current or voltage measurement function.
Parameter
0|OFF|1|ON, *RST OFF
Enables datalog current measurements: SENS:ELOG:FUNC:CURR ON
SENSe:ELOG:FUNCtion:CURRent:MINMax 0|OFF|1|ON
SENSe:ELOG:FUNCtion:CURRent:MINMax?
SENSe:ELOG:FUNCtion:VOLTage:MINMax 0|OFF|1|ON
SENSe:ELOG:FUNCtion:VOLTage:MINMax?
Enables or disables logging of the minimum and maximum current or voltage values.
Parameter
0|OFF|1|ON, *RST OFF
Enables MIN/MAX logging values: SENS:ELOG:FUNC:VOLT:MINM ON
SENSe:ELOG:PERiod <value>|MIN|MAX
SENSe:ELOG:PERiod? [MIN|MAX]
Sets the integration time of an Elog measurement.
Keysight RP7900 Series Operating and Service Guide
Typical Return
(none)
Typical Return
0 or 1
Typical Return
0 or 1