STATus:PRESet
Presets all Enable, PTR, and NTR registers.
Operation register
STAT:OPER:ENAB
STAT:QUES<1|2>:ENAB
STAT:OPER:NTR
STAT:QUES<1|2>:NTR
STAT:OPER:PTR
STAT:QUES<1|2>:PTR
Parameter
(none)
Preset the Operation and Questionable registers: STAT:PRES
STATus:QUEStionable<1|2>[:EVENt]?
Queries the
event register
stores (latches) all events that are passed by the Operation NTR and/or PTR filter. Reading the
Questionable Status Event register clears it.
Parameter
(none)
Read questionable status event register #1: STAT:QUES1?
The value returned is the binary-weighted sum of all enabled bits in the register.  For example, with
l
bit 2 (value 4) and bit 4 (value 16) set, the query returns +20.
*RST has no effect on this register.
l
STATus:QUEStionable<1|2>:CONDition?
condition register
Queries the
holds the live (unlatched) operational status of the instrument. Reading the Questionable Status
Condition register does not clear it.
Parameter
(none)
Read questionable status condition register #1: STAT:QUES1:COND?
The value returned is the binary-weighted sum of all enabled bits in the register.  For example, with
l
bit 2 (value 4) and bit 4 (value 16) set, the query returns +20.
The condition register bits reflect the current condition. If a condition goes away, the cor-
l
responding bit is cleared.
*RST clears this register, other than those bits where the condition still exists after *RST.
l
Keysight RP7900 Series Operating and Service Guide
Questionable register
all defined bits are disabled
all defined bits are disabled
all defined bits are enabled
for the
Questionable Status
Questionable Status
for the
5 SCPI Programming Reference
Preset setting
Typical Return
(none)
group. This is a read-only register, which
Typical Return
<bit value>
group. This is a read-only register, which
Typical Return
<bit value>
235

Publicité

loading