1 Quick Reference
Common Characteristic
Acoustic statement
(European Machinery Directive)
Digital Port
Max voltage rating:
Pins 1 & 2 as FLT:
Pins 1-7 as outputs:
Pins 1-7 as inputs:
Pin 8:
Regulatory Compliance
EMC:
Safety:
AC Input
Phase and range:
Frequency:
Input VA:
Input current per phase
200 VAC input:
400 VAC input:
Efficiency at full power:
Power Factor:
Output Terminal Isolation:
For 20 VDC models:
For 80 & 160 VDC models:
38
Acoustic noise emission
LpA 79 dB at Operator position
LpA 73 dB at Bystander position
LpA 61.4 at Idle fan speed
Normal operation mode per ISO 7779
+16.5 VDC/- 5 VDC between pins
Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA
Typical high-level leakage current = 1 mA @ 16.5 VDC
Maximum low-level output voltage = 0.5 V @ 4 mA;
1 V @ 50 mA; 1.75 V @ 100 mA
Maximum low-level sink current = 100 mA
Typical high-level leakage current = 0.8 mA @ 16.5 VDC
Maximum low-level input voltage = 0.8 V
Minimum high-level input voltage = 2 V
Typical low-level current = 2 mA @ 0 V (internal 2.2k pull-up)
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Pin 8 is common (internally connected to chassis ground)
Complies with European EMC Directive for test and measurement products
Complies with Australian standard and carries C-Tick mark
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Complies with European Low Voltage Directive and carries the CE mark.
Conforms to US and Canadian safety regulations.
Declarations of Conformity for this product may be downloaded from the Web. Go to
http://www.keysight.com/go/conformity
3 phase; 200 VAC nominal ±10%, and 208 VAC nominal ±10%
3 phase; 400 VAC nominal ±15%, and 480 VAC nominal ±10%
50/60 Hz
RP7931A, RP7932A, RP7941A, RP7942A, RP7936A: 6.5 kVA
RP7933A, RP7935A, RP7936A, RP7943A, RP7945A, RP7946A: 11.5 kVA
RP7931A, RP7932A: 17.3 A; RP7933A, RP7935A, RP7936A; 35 A
RP7941A, RP7942A: 8.66 A; RP7943A, RP7945A, RP7946A: 17.3 A
RP7931A, RP7932A, RP7941A, RP7942A: 84%
RP7933A, RP7935A, RP7936A, RP7943A, RP7945A, RP7946A: 85%
0.99 @ nominal input and rated power
During a 1-cycle line dropout the unit may reboot. The output will remain off after reboot until the
operator reinstates the previous settings, either by the front panel controls or using a computer
program. This behavior is consistent with safe operating procedures.
No output terminal may be more than ±60 VDC from any other terminal or chassis ground.
No output terminal may be more than ±240 VDC from any other terminal or chassis ground.
All Models
and click on "Declarations of Conformity."
Keysight RP7900 Series Operating and Service Guide