Parameter
0 – 262.144, *RST 1 ms
Programs a dwell list. The list contains 3 steps: LIST:DWEL 0.2,0.8,1.6
[SOURce:]LIST:CURRent:POINts?
[SOURce:]LIST:DWELl:POINts?
[SOURce:]LIST:VOLTage:POINts?
[SOURce:]LIST:TOUTput:BOSTep:POINts?
[SOURce:]LIST:TOUTput:EOSTep:POINts?
Returns the number of list points. Points are the same as steps. The queries do not return the point
values.
Parameter
(none)
Returns the number of points in the dwell list: LIST:DWEL:POIN?
[SOURce:]LIST:STEP ONCE|AUTO
[SOURce:]LIST:STEP?
Specifies how the list responds to triggers. ONCE causes the output to remain at the present step until
a trigger advances it to the next step. Triggers that arrive during the dwell time are ignored. AUTO
causes the output to automatically advance to each step, after the receipt of an initial starting trigger.
Steps are paced by the dwell list. As each dwell time elapses, the next step is immediately output.
Parameter
ONCE|AUTO, *RST AUTO
Specifies the list steps to be paced by trigger signals: LIST:STEP ONCE
[SOURce:]LIST:TERMinate:LAST 0|OFF|1|ON
[SOURce:]LIST:TERMinate:LAST?
Determines the output value when the list terminates. When ON (1), the output voltage or current
remains at the last list step. The value of the last voltage or current list step becomes the IMMediate
value when the list completes. When OFF (0), and also when the list is aborted, the output returns to
the settings that were in effect before the list started.
Parameter
0|OFF|1|ON, *RST OFF
Terminate with the output at the last step value: LIST:TERM:LAST ON
Keysight RP7900 Series Operating and Service Guide
Typical Return
<list value 1>,<list value 2>,<list value 3>
Typical Return
<points>
Typical Return
ONCE or AUTO
Typical Return
0 or 1
5 SCPI Programming Reference
217