Front Panel Menu Reference
Select Transient\Arb\Config.
If Arb points have been imported or programmed using
the SCPI command, the Points field displays the
number of points in the Arb.
View the amplitude of any Arb point by entering the
point number in the Point # field. The Level field
displays the amplitude.
Specify how many times you want the Arb to repeat
Depending on your application, specify how many times you want the Arb to repeat. Sending the
INFinity parameter in the SCPI command makes the Arb repeat indefinitely. At reset, the Arb count is
set to 1.
Front Panel Menu Reference
Select Transient\Arb\Repeat.
Enter the number of list repetitions (2) and press
Select.
Specify how you want the Arb to end
Specify the output state after the Arb has completed. There are two choices: the output returns to the
state it was in before the Arb started, or the output remains at the values of the last Arb point.
Front Panel Menu Reference
Select Transient\Arb\Terminate.
Select either Return to Start, or Stop at Last Step and
press Select.
Keysight RP7900 Series Operating and Service Guide
4 Using the Regenerative Power System
SCPI Command
To program 10 points in a current Arb:
ARB:CURR:CDW 1,2,2,3,4,4,3,2,2,1
To query the number of Arb points:
ARB:CURR:CDW:POIN?
To query the Arb point values:
ARB:CURR:CDW?
SCPI Command
To program the Arb to repeat twice:
ARB:COUN 2
SCPI Command
To return the output to the pre-Arb state:
ARB:TERM:LAST OFF
To keep the output at the Arb end point:
ARB:TERM:LAST ON