Miltenyi Biotec MACSQuant X Mode D'emploi page 86

Publicité

Parameter
Minimum sample
volume
Scatter resolution
Sample flow rate
Maximal event rate
Lasers
Fluorescence sensitivity
(MESF)
Absolute cell count
performance
Sample carryover
Emission detectors
Table 9.2: Operation details of the MACSQuant X
9
84
Specification
1 μL (20 µL recommended for volumetric counting applications)
Scatter performance is optimized for resolving human peripheral blood lymphocytes,
monocytes, and granulocytes
25-100 μl/min plus automated flow rate to maintain 500, 1000, or 2000 events/second
Up to 15,000 events/second
405 nm, 40 mW
488 nm, 30 mW
640 nm, 20 mW
FITC < 200; PE < 100; APC < 150
Accuracy +/- 7%; reproducibility (CV) <5%
≤ 0.1% in Standard mode
Channel
FSC
SSC
V1
V2
B1
B2
B3
B4
R1
R2
Filter
488/10 nm
488/10 nm
450/50 nm
525/50 nm
525/50 nm
585/40 nm
655-730 nm
750 nm LP
655-730 nm
750 nm LP