2.6
Assembly
2.6.1
Install the fluid bottles
Operating the instrument requires Running Buffer, Washing Solution, and Storage Solution. One bottle is
required for waste collection. Always operate the instrument with ready-to-use MACS Buffers and solutions
only. The bottles, closures, tubing and fluid sensor cables are color-coded for easier handling.
1 Install only one fluid bottle at a time. Place the new bottle into the orange bottle holder. Please note the
corresponding color coding.
2 Unscrew the lid of the fluid bottle and replace it with the appropriate bottle closure delivered with the
instrument.
2
Figure 2.1: Connect Waste and Storage Solution bottle
52