Miltenyi Biotec MACSQuant X Mode D'emploi page 60

Publicité

Needle priming time: defines the priming time when a new experiment is started to avoid that the needle runs dry
l
(20 seconds are recommended)
3 Go to Shutdown default behavior.
4 Select Instrument off or Data analysis mode as the default behavior for all users.
3.4
Long-term storage
The instrument must be run every two weeks in order to prevent clogging. If the instrument is not in use, turn
on the instrument every two weeks and switch it into acquisition mode. During the priming procedure, the
fluidics will be flushed.
1 Switch on the instrument.
2 To prime the instrument (power up the lasers, activate fluidics), click on the Main instrument control
button in the upper right corner of the software and select Acquisition mode.
3 Right-click on the Rinse button and select Flush.
3
4 Shut down the instrument.
58