4 Connect the 4-end fluid sensor cable to the port labeled Bottle Sensor at the back of the instrument
(Figure 1.2) and fasten securely.
5 Note the color coding and connect each fluid sensor cable to the respective bottle closure.
6 Connect a hydrophobic air filter to the appropriate connectors on the bottle closures to keep buffers sterile.
Avoid any contact with liquids, as this may cause clogging of the filter.