Télécharger Imprimer la page

GETINGE Maquet Axcel AXL 5001 Manuel D'utilisation page 47

Masquer les pouces Voir aussi pour Maquet Axcel AXL 5001:

Publicité

Eclairage Interventionnel
Surgical Light System
Lámparas de Intervención
MAQUET SAS project No.
Date of delivery or of dispatch FOB
Date of installation
Date of commissioning
After Sales Department
Stamp
Installer (print name)
Customer (print name)
Inspections to be carried out before placing in service
1 Ground continuity: from bulb cover hinge to cupola(s) to suspension tube ground screw
2 Operation of light bulbs
3 Maximum voltage at bulb-holder
4 Verticality of suspension tube
5 Balancing of intermediate arm (main cupola)
Balancing of suspension arm (satellite cupola)
6 Click-in locking of sterilizable handles
7 General aspect and cleaning
Manuel d'utilisation / User's manual / Manual del Usuario
INSPECTION SHEET
(copy to be kept by the customer)
Customer identification
Signature
Signature
V (
)
RMS
V
23,5
EFF
AC+DC
0125104
Serial No.
Product Ref.
Operating theatre ref.
Operating room ref.
CUSTOMER'S ACCEPTANCE :
Customer (print name) :
Office :
Easy handiness :
Correct stability :
Date
Date
q
yes
q
no
q
yes
q
no
Inspection carried out
q
q
q
q
q
q
q
q
47