Table 202 - Guidance and manufacturer's declaration - electromagnetic immunity
AXCEL surgical lights are intended for use in the electromagnetic environment specified below. AXCEL customers or users
should ensure that they are used in such an environment..
Immunity test
Electrostatic
± 6 kV contact
discharge
± 8 kV air
(ESD) IEC 61000-4-2
Electrical fast
± 2 kV for power
transient / burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV differential
mode
IEC 61000-4-5
± 2 kV common
mode
Voltage dips, short
< 5% U
interruptions and
(dip > 95% of U
voltage variations on
for 0.5 cycles
power supply lines
40% U
IEC 61000-4-11
(dip = 60% of U
for 5 cycles
70% U
(dip = 30% of U
for 25 cycles
< 5% U
(dip = 95% of U
for 5 cycles
Power frequency
3 A
(50/60 Hertz)
magnetic fields
IEC 61000-4-8
Note: U
is the voltage of the AC supply network before applying the test level.
T
Manuel d'utilisation / User's manual / Manual del Usuario
IEC 60601
Compliance
test level
level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV differential
mode
± 2 kV common
mode
< 5% U
T
T
)
(dip > 95% of U
T
for 0.5 cycles
40% U
T
T
)
(dip = 60% of U
T
for 5 cycles
70% U
T
T
)
(dip = 30% of U
T
for 25 cycles
< 5% U
T
T
)
(dip = 95% of U
T
for 5 cycles
3 A
Electromagnetic environment — guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
)
commercial or hospital environment. If the AXCEL
T
requires continuous operation during mains
interruptions, it is recommended that it be powered from
an uninterruptible power supply.
)
T
)
T
)
T
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
0125104
31