Lifting the stretcher
To bring back the stretcher back to the standard height, starting from any of the previous configurations, the operators must coordinate while simultaneously lifting the front
and rear of the stretcher, ensuring the proper alignment of the stretcher and the incubator. Do not carry out these operations until you have checked that the patient has been
properly positioned and immobilized inside the incubator.
• At the rear, grab the part of the frame near the leg unlocking handles.
• At the front, similarly grab the frame above the loading trolley.
• Raise the assembly until the locking mechanisms are properly engaged.
Only use the frame for lifting.
Incubator supports:
The trolley is equipped with two supports on which the incubator can be installed.
The maximum usable width of each support is 400mm, while the distance is adjustable in the range 360-800mm.
min 360mm
max 800mm
11. CLEANING AND MAINTENANCE
Spencer Italia S.r.l. declines all responsibility for any direct or indirect damage which is the consequence of improper use of the product and spare parts and/or in any
case of any repair carried out by a person other than the Manufacturer, who uses internal and external technicians authorised to do so; moreover, doing so will invalidate
the warranty.
• The operator must wear suitable personal protective equipment, such as gloves, goggles, etc. during all checking, maintenance and cleaning operations.
• Establish a maintenance schedule, periodic inspections and extend the average life span, if foreseen by the Manufacturer in the Instructions for Use, identifying a reference
person who meets the basic requirements set forth in the Instructions for Use.
• The frequency of inspections is determined by factors such as legal requirements, type of use, frequency of use, and environmental conditions during use and storage.
• Repairs of products manufactured by Spencer Italia S.r.l. must be carried out by the Manufacturer, who shall make use of specialised internal or external technicians
who, using original spare parts, shall provide quality repair service in strict compliance with the technical specifications indicated by the Manufacturer. Spencer Italia S.r.l.
declines any responsibility for any direct or indirect damage which is a consequence of improper use of spare parts and/or any repair work carried out by unauthorised
parties.
• Use only original or Spencer Italia S.r.l. approved components/spare parts and/or accessories in order to carry out all operations without causing alterations or modifications
to the product.
• All maintenance and overhaul activities must be recorded and documented with the relevant technical operation reports. This documentation must be kept for at least 10
years after the end of the product's life and must be made available to the competent authorities and/or the Manufacturer when requested.
• Cleaning, provided for reusable products, must be carried out in accordance with the Manufacturer's instructions in the Instructions for use in order to avoid the risk of
cross-infection due to the presence of body fluids and/or residues.
• The product and all its components must be washed and left to dry completely before storage.
• If the product requires lubrication, this must be done after cleaning and complete drying.
• Document maintenance on appropriate forms. .
11.1 CLEANING
Failure to carry out the correct cleaning operations could increase the risk of cross-infection due to presence of body fluids and/or residues.
The operator must wear suitable personal protective equipment, such as gloves, goggles, etc. during all checking and cleaning operations.
Metal parts exposed to external agents undergo surface treatments and/or coating in order to obtain better resistance. Clean the exposed parts with water and delicate soap.
Never use solvents or stain removers.
Do not use cleaning agents containing sodium hypochlorite as corrosion of components may occur.
Rinse thoroughly with lukewarm water, making sure you have removed all traces of soap, which may deteriorate or compromise conditions and durability. Avoid using high
pressure water, as it penetrates the joints and removes lubricant, creating the risk of corrosion on components. Let dry completely before storing. Drying after washing or after
use in a wet environment must be natural and not forced. Do not use flames or other direct heat sources.
If disinfecting, use products that do not have a solvent or corrosive action on materials constituting the device. Be sure to take all precautions to ensure that there is no risk of
cross-infection or contamination of patients and operators.
11.2 ROUTINE MAINTENANCE
Establish a maintenance program and periodic checks, identifying a designated reference person. The person entrusted with maintenance of the device must ensure the
basic requirements envisaged by the Manufacturer within these instructions for use.
All routine and special maintenance activities and all general overhauls must be recorded and documented with the relevant technical intervention reports. This
documentation must be kept for at least 10 years after the end of the device's life and must be made available to the competent authorities and/or the Manufacturer
when requested.
The operator must wear suitable personal protective equipment, such as gloves, goggles, etc. during all checking, maintenance and cleaning operations.
Checks to be carried out before and after each use and in any case every 3 months are as follows:
• General functionality of the device
• Cleanliness of the device (remember that the failure of cleaning may cause the risk of cross infections)
• Correct tightening of all nuts, bolts and screws
• Absence of holes, tears or abrasions on the entire structure
• None of the tubes or metal sheets present bends or cracks
• Welded areas are intact, without any cracks or breaks
• Moving parts, wheels, levers, handles are intact and work properly
After having identified the suitable position in relation to the incubator in use, it is possible to lock the sup-
ports by tightening the locking screws until they come to a stop, then proceed with the drilling necessary
to fix the incubator, which must pass through the rectangular-section tube.
400 mm
The choice of the number of holes, their positioning and the type of screws to be used is the responsi-
bility of the installer, who is responsible for assessing compliance of the fixing in accordance with the
regulations in force.