8.
Maintenance
To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting list
►
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Cause
Product not running
No battery
Battery not charged
Battery defective
Product in secured OFF posi-
tion
Product defective
Product is getting too hot Excessive use
Reprocessing/maintenance
carried out incorrectly
Fall damage, product defec-
tive
Blunt tool
Insufficient power
Product defective
Blunt tool
Loud running noise
Defective product drive/ball
bearings
Lid cannot be
Lid not compatible
attached/removed
Lid deformed/defective
Connection to product
deformed/defective
Battery cannot be
Battery not compatible
inserted/removed
Battery deformed/defective
Battery compartment on
product deformed/defective
Detection
No battery in the battery compartment Insert battery.
No beep when inserting the battery
No signal tone when inserting the bat-
tery
Safety catch is in OFF position
Product not running
Product is heating up
Product heating
Product heating
Tool and product are heating up
Inadequate product power
Large amount of heat generated after a
short time
Tool blades worn out
Loud, perceptible noise during opera-
tion
Lid does not click into place
Lid is difficult or cannot be
mounted/removed
Lid is hard or impossible to fit/remove
Battery cannot be completely inserted
into the battery compartment
Battery is is difficult to insert/remove
or cannot be inserted/removed
Battery is difficult or cannot be
mounted/removed
Remedy
Charge battery in charger.
Have the battery repaired by the manufacturer.
Turn the safety catch to the ON position.
Have the product repaired by the manufacturer.
Observe instructions for use (nominal operating
mode).
Follow operating instructions (preparation, care).
Preventative: lubricate product before every steril-
ization.
Have the product repaired by the manufacturer.
Change tool.
Follow operating instructions (preparation, care).
Preventive measure: Lubricate product before
every sterilization.
Have the product repaired by the manufacturer.
Follow operating instructions (operating mode).
Have the manufacturer repair the product
Replace tool.
Observe instructions for use (processing, care).
Preventive measure: Lubricate product before
every sterilization.
Have the manufacturer repair the product
Use the compatible lid for GA331.
Have the manufacturer repair the lid.
Have the product repaired by the manufacturer.
Use the compatible battery for GA331.
Have the manufacturer repair the battery.
Have the product repaired by the manufacturer