Télécharger Imprimer la page

KERR ENDODONTICS elements IC Mode D'emploi page 9

Système d'obturation sans fil

Publicité

Problem
The light on charger
fails to turn from orange
to blue
Battery becomes
exhausted quickly
LED on the Charging Base
blinks every two seconds
(Or within seconds)
Device shows 'ERROR
CODE'
(ERROR CODE #1˜5)
SPECIFICATIONS
• Battery: 3.6 V, 1900 mAh Li-Ion, 6.8 Wh
• Adapter: Input: 100-240V~, 50/60 HZ, Output 9V 2A
• Downpack: ~161mm x 33mm x 30mm (L x W x H), ~100 g
• Backfill: ~206mm x 33mm x 30mm (L x W x H), ~172 g
• Device: IEC 60601-1 Edition 3.1, Class I, patient contact,
Type B Applied part
• Mode of Operation: Continuous Operation
• Wireless Charging Frequency:
– Dual charger: 143.96kHz
– Single charger (Backfill): 146.7kHz
– Single charger (Downpack): 147.1kHz
Storage and Transport Conditions:
50°C (122°F)
-20°C (-4°F)
Temperature
10%
Atmospheric Pressure
1400 hPa
500 hPa
Atmospheric Pressure
WARRANTY
Kerr Endodontics warrants the system (excluding batteries) to be free
from defects in materials or workmanship for period of 2 years from the
original date of purchase. The batteries are warrantied for a period of 6
months from the original date of purchase. If the system shows any defect
within the warranty period that are not excluded from this warranty, Kerr
Endodontics shall, at its sole discretion, either replace or repair the device
using suitable new or reconditioned parts. In the case other parts are used
which constitutes an improvement, Kerr Endodontics may, at its discretion,
charge the customer for the additional cost of these parts. If the warranty
claim provides to be justified, the product will be returned to the user freight
prepaid. Warranty claims other than those indicated herein, are expressly
excluded.
EXCLUSIONS
Damage and defects caused by the following conditions are not covered by
the warranty:
077-0901_REVA_ELEMENTS_IC_EU_IFU_BOOK_P24.indb 7
Cause
Solution
Battery may be
Wait at least 4 hours.
near end-of-life
If unit fails to reach full
charge, replace battery.
Only use Kerr Endodontics
battery.
Battery may be
Replace battery. Only
near end-of-life
use Kerr Endodontics
battery. Note: On the
Backfill device, battery has
enough power
for at least two procedures.
It is recommended to
place the
device on the Charging
Base between uses.
Incorrect charging
Remove device from
the Charging Base and
recharge. If the problem
continues, then unplug the
transformer from the dock
and plug it again.
One or more
Contact Customer Care to
internal
get the device repaired.
components are not
working properly
Conditions for use:
35°C (95°F)
10°C (50°F)
Temperature
75%
90%
30%
Atmospheric Pressure
1013 hPa
697 hPa
Atmospheric Pressure
0-10,000 ft.
Altitude Limitation
• Improper handling/disassembly/modifying, neglect, or failure to
operate the unit in compliance with the instructions given in this
manual.
• Force majeure or any other condition that is beyond the control of Kerr
Endodontics.
• Damage caused by customer misuse or uses other than those specified.
DISCLAIMER
For safety reasons, this product should be used with accessories
manufactured and sold by Kerr Endodontics. Any use of non-authorized
accessories or not following any of the instructions for use is done so at the
operator's risk and voids the warranty. Kerr Endodontics does not assume
any responsibility for incorrect diagnosis due to operator error or equipment
malfunction.
PART NUMBERS
Parts
elements™ IC Obturation System
Downpack Unit
Backfill Unit
Dual Charger
Single Charger
Transformer
Power Cord
Battery
Heat Pluggers
Medium Large (ML)
Medium (M)
Fine Medium (FM)
Fine (F)
Extra-Fine (XF)
elements Cartridge
Gutta Percha, Medium Body, 23 Gauge - SILVER
Gutta Percha, Heavy Body, 23 Gauge - SILVER
Gutta Percha, Light Body, 25 Gauge - SILVER
Gutta Percha, Medium Body, 23 Gauge - GOLD
Gutta Percha, Heavy Body, 23 Gauge - GOLD
Gutta Percha, Light Body, 25 Gauge - GOLD
Gutta Percha SILVER
NOTE:
Elements gutta-percha cartridges availability might be different depending
on different regions of the world.
*X designates plug type
Type B
Type F
Type G
Type I
Type N
Reorder Code
973-0600-TYPEX*
973-0602-TYPEX*
973-0604-TYPEX*
973-0610
973-0612
973-0615
973-0616-TYPEX*
973-0620
Reorder Code
952-0007
952-0006
952-0005
952-0004
952-0031
Reorder Code
972-1002
972-1005
972-1003
972-2500
972-2502
972-2501
Gutta Percha GOLD
7
6/13/19 6:21 PM

Publicité

loading

Ce manuel est également adapté pour:

973-0600-typex