4.3 Selecting sterilisation programs
4.3
01 Select the "Sterilisation cycles" menu
02 Alternatively, directly select "Last cycle run" ando go
to step 04
03 Select the desired cycle
04 Screen with all the parameters of the selected cycle
05 (Optional) You can open them door to insert other instruments
06 (Optional) Select the programmed start button to set
the start time
07 (Optional) Select the increment drying button to increase
the cycle drying time
08 Start the cycle
09 (If active) enter the user PIN.
On confirming, the cycle will start
10 You can use the green bar to monitor the cycle progress
11 When the cycle finishes, press the confirm button to return
to the Home screen
12 You can stop the cycle manually by holding done the
"Stop cycle" buttonh for 3 seconds. Wait for the machine
to be put in a safe condition and press the confirm button (13)
for 2 seconds and then press Home to return to the home screen.
*Opzioni 06 e 07 disponibili solo nella versione Premium
*Options 06 and 07 are only available in the Premium version
4.3.1 Sterilisation program types
When the machine is cold, the preheating time is approximately 10 minutes. Selecting the "Programs" submenu
accesses the list of sterilisation cycles that the autoclave can run:
B CYCLES
- 134°C Standard Cycle: 4 min. of sterilisation plus the drying phase
- 134°C Prion Cycle: 18 min. of sterilisation plus the drying phase
- 134°C B-Flash Cycle: 3,5 min. of sterilisation plus the drying phase
- 121°C Standard Cycle: 16 min. of sterilisation plus the drying phase
S CYCLES
- 134°C Universal Cycle: 3,5 min. of sterilisation plus the drying phase
- 121°C Softprogramm Cycle: 15,5 min. of sterilisation plus the drying phase
SPECIAL CYCLES
- Drying cycles: drying phase only
NOTE: For each available sterilisation cycles, the manufacturer must perform the Works Tests in accordance with EN
13060 (par. 7.3). Any available cycles for which the manufacturer has not performed the Works Tests must undergo
Installation Tests in accordance with EN 13060 par. 7.4.