2.2.7 USB port
This port can be used by medical staff for data download .
2.2.8 Serial port - RS232
This port is only for connecting the FARO printer.
Use the dedicated serial cable supplied with the Faro printer.
2.2.9 Touch screen display
This manages the autoclave fully through a menu system.
2.2.10 LAN connection
This port is only for connecting a LAN.
2.2.11 Safety devices
The steriliser is equipped with the following safety devices:
- Mains fuses
These protect the equipment from faults in the heating elements by disconnecting the electric power supply.
- Electronic circuit protection fuses
These prevent faults in the transformer primary circuit and low-voltage loads by disconnecting one or more low-
voltage circuits.
- Safety valve
This prevents any overpressure in the sterilisation chamber by releasing steam to restore a safe pressure. It opens
at a pressure of 350kPa (±10%).
- Steam generator thermal switch
This stops the steam generator from overheating by disconnecting its electrical power supply. It trips at a tempe-
rature of 220°C ±7°C.
- Chamber thermal switch
This protects the pressure vessel heating element from overheating by disconnecting its electric power supply. It
trips at a temperature of 150°C ±4°C.
- Door lock mechanism safety microswitch
This provides feedback when the door and the door lock system are not in the correct position or when the door
lock mechanism is not working correctly.
- Pressure switch
This prevents the door from being opened accidentally during the program when the machine is pressurised (even
in the event of a blackout).
- Overflow system
This prevents water from flowing out onto live parts if the level sensor does not work by draining excess water
to the outside.
- Built-in sterilisation process assessment system
This uses a microprocessor to continuously monitor the sterilisation process parameters. In the event of a fault, it
immediately stops the program and generates alarms.
- Steriliser operation monitoring
When the machine is powered, this monitors all of its important parameters in real time. In the event of a fault, it
generates alarm messages and, if necessary, stops the cycle.
TYPE TESTS cARRIED OUT ON NISEA SERIES AUTOcLAVES
Type test carried out in accordance with the EN13060 standard Type B cycles
Pressure dynamic in sterilisation chamber
Air leakage
Empty chamber
Solid load - Double wrapping
Small porous load - Double wrapping
Full porous load - Double wrapping
Narrow lumen element
Drying, solid load - Double wrapping
Drying, full porous load - Double wrapping