Code
ER 100
ER 101
ER 110
ER 120
ER 121
ER 130
ER 140
Code
ER 141
ER 150
ER 151
ER 152
ER 153
ER 154
Code
ER 160
ER 170
ER 180
ER 200
ER 201
ER 202
ER 203
ER 204
72
AL 001
Possible cause
There is a leak in the machine plumbing circuit or a
malfunction in the vacuum pump.
There is a leak in the machine plumbing circuit.
There is a leak in the machine plumbing circuit or a
malfunction in the vacuum pump.
There is a leak in the machine plumbing circuit or a
malfunction in the water injection pump.
There is a leak in the machine plumbing circuit or a
malfunction in the water injection pump.
There is a malfunction in the machine discharge system.
There is a leak in the machine plumbing circuit or a
Possible cause
malfunction in the water injection pump.
There is a leak in the machine plumbing circuit or a
malfunction in the water injection pump.
There is a malfunction in the machine discharge system
or a chamber loading error.
There is a malfunction in the sterilisation condition
holding system or a chamber loading error.
There is a malfunction in the sterilisation condition
holding/control system
There is a malfunction in the sterilisation condition
holding/control system
There is a malfunction in the sterilisation condition
holding/control system
Possible cause
There is a malfunction in the machine discharge system
There is a malfunction in the machine drying system
There is a malfunction in the pressure balancing system
There is a leak in the machine plumbing circuit or a
malfunction in the vacuum pump.
There is a leak in the machine plumbing circuit.
There is a leak in the machine plumbing circuit.
A Vacuum Test is running in chamber conditions other
than ambient temperature.
There is a malfunction in the pressure balancing system
The reset button was pressed
Possible solution
Reset the machine.
Check that the door seal is clean and positioned correctly, then repeat
the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that the filter in the chamber is clean and positioned correctly.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check the amount of load in the chamber and repeat the selected
program.
If the problem persists, contact technical support.
Reset the machine.
Possible solution
Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check:
a) the tray support position and the tray positions in the chamber;
b) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check:
a) that there is water in the feed tank;
b) the amount of load in the chamber.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check:
a) the tray support position and the tray positions in the chamber;
b) that there is water in the feed tank;
c) the amount of load in the chamber.
Possible solution
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check the amount of load in the chamber and then repeat the selected
program.
If the problem persists, contact technical support.
Reset the machine.
Check the amount of load in the chamber and the Bio-X filter
condition, then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check:
a) that the door seal is clean and positioned correctly;
b) that the filter in the chamber is clean and positioned correctly.
Then repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.
Check that the door seal is clean and positioned correctly, then repeat
the selected program.
If the problem persists, contact technical support.
Reset the machine.
Repeat the Vacuum Test, making sure that the chamber is at ambient
temperature.
If the problem persists, contact technical support.
Reset the machine.
Repeat the selected program.
If the problem persists, contact technical support.
Reset the machine.