Cycle error
Cycle progress
Sterilisation
complete
Feed tank
maximum level
Feed tank
intermediate level
Feed tank
minimum level
Water quality
Water quality bar
4.2
4.2
• Home
4.2 Menu architecture
Home
●
●
●
●
●
●
Data output
●
Open door
●
Water filling
●
* In the Premium version only
54
bar
Menu architecture
Menu architecture
Sterilisation programs
•
Last sterilisation program run*
Sterilisation programs
•
Test programs
Last sterilisation program run*
•
Last test program run*
Test programs
•
Settings
Last test program run*
•
Settings
User management*
•
User management*
Date and time
●
•
Date and time
Printer
●
•
Printer
Language
●
•
Language
Measurement unit*
●
•
Measurement unit*
●
Water parameters*
•
Water parameters*
●
Connectivity*
•
Connectivity*
●
Service
•
Service
●
• Data output
USB
•
Usb
●
Printer
•
Printer
●
Info
•
Info
●
Open door
•
Water filling
•
* In the Premium version only
Cannot be selected.
Appears when the cycle has been stopped by an error or
manually by the operator.
The cycle progress bar advances according to the following
phases:
1.
Preheating phase
2.
Vacuum phase 1
3.
Pressure ramp-up phase 1
4.
Vacuum phase 2
5.
Pressure ramp-up phase 2
6.
Vacuum phase 3
7.
Pre-sterilisation phase
8.
Balancing phase
9.
Sterilisation phase
10. Discharge phase
11. Drying phase
12. Pressure balancing
Indication icon only.
Cannot be selected.
Appears when the sterilisation phase is complete (not the
cycle).
Allows the user to stop the cycle to remove the instruments,
skipping the drying phase
Appears when the water in the feed tank reaches the maximum
level. Water filling stops automatically.
Appears when the water in the feed tank is between the
minimum and maximum levels.
Appears when the water in the feed tank is close to the
minimum level.
Indication icon only.
Cannot be selected.
Shows the water quality bar
The cursor position indicates the water quality.
Green: quality OK
Yellow: critical quality area message generated
Red: quality not OK message generated and machine stops