13 Technical Data

Input data
Sensor type
1)
Probe length
Measuring range
Max. speed of change of the fluid level
2)
Repeatability
Switching point accuracy
Temperature (optional)
Sensor type
Measuring range
Accuracy
Reaction time (t
)
90
Output data
Analogue output (optional)
With 1 or 2 SP selectable
With 4 SP (only with temperature sensor)
Switching outputs
Version
Assignment
Switching current
Switching cycles
Ambient conditions:
Operating temperature range
Storage temperature range
Fluid temperature range
- mark
Vibration resistance according to
DIN EN 60068-2-6 (0 .. 500 Hz)
Shock resistance according to
DIN EN 60068-2-29 (11 ms)
Protection class to IEC 60529
Version 2018/07/06
3)
HYDAC ELECTRONIC GMBH
Magnetostrictive Fluid Level Sensor
250; 280; 370; 410; 520; 730 mm
178; 208; 298; 338; 448; 658 mm
No restrictions
≤ ± 1 % FS
≤ ± 1 % FS
Semi-conductor sensor
-25 .. +100 °C
± 1.5 °C
≤100s
ohmic resistance ≤ 500 Ω
4 .. 20 mA
0 .. 10 V
ohmic resistance ≥ 1 kΩ
Corresponds to measuring range selected
0 .. 10 V
ohmic resistance ≥ 1 kΩ
Corresponds to measuring range selected
Transistor output PNP
Programmable as N/O or N/C
On version with temperature measurement
user-selectable temperature or fluid level
1 or 2 SP: max. 1.2 A per output
4 SP:
max. 0.25 A per output
> 100 million
-25 .. +80 °C
-40 .. +80 °C
-40 .. +120 °C / -25 .. +120 °C
EN 61000-6-1 / 2 / 3 / 4
≤ 2 g
≤ 20 g
IP 67