M.blue
®
APPLICATION OF M.blue plus Instruments
The M.blue plus Instruments may only be used
by trained specialist personnel. The selected
pressure level of M.blue can be determined,
adjusted and monitored using the M.blue plus
Instruments. The M.blue plus Compass is
used for localising and reading the adjustable
gravitational unit of M.blue.
Fig. 7: M.blue plus Compass
a) open b) closed
1. Scale Ring 2. Float gauge compass
With the M.blue plus Adjustment Ring the
opening pressure of M.blue can be variably
adjusted from 0 to 40 cmH
Fig. 8: M.blue plus Adjustment Ring
The opening pressure for the adjustable grav-
itational unit can be changed before or after
implantation. It is preset by the manufacturer to
20 cmH
O. Please carry out the following steps
2
to adjust the valve:
O.
2
INSTRUCTIONS FOR USE | EN
1. Localisation
When the M.blue plus Compass is opened, a
circular template becomes visible with which
you can localise the valve in the patient's head
using your index finger (fig. 9).
Fig.
9:
Localising
M.blue plus Compass
The direction markings on the template show
the flow direction.
2. Test procedure
In order to determine the selected pressure
level, the compass is then closed again. The
float gauge should now be centred by moving
the instrument in the designated circular mark-
ing (fig. 10). Once the float gauge is centred,
the currently set opening pressure of the gravi-
tational unit can be read off via the line mark on
the float gauge (fig. 10).
Fig. 10: Determination of the pressure level with the
M.blue plus Compass
the
valve
using
the
21