7.1

Instructions for use

Before using the unit-mounted MASTER FLUX PLUS version, lock the wheels that have
brakes on them by pressing the tabs downwards.
If it becomes necessary to move the appliance unit (e.g. from one room to another), please
unblock the wheels and push the device gently using the handles (see Paragraph 2.3)
Avoid shocks and sudden movements that could cause accidental falls.
Move and position the MASTER FLUX PLUS only on non-inclined and
smooth surfaces.
Make sure the cylinders are stable and secured correctly.
Close the cylinders before moving the device.
Do not use the handles to lift the device.
Do not move the device using the handles.
Do not expose the controls to involuntary knocks.
When using MASTER FLUX PLUS in its wall-mounted version, it is possible to move and
position the flowmeter box as required by gripping the relative handle (Fig 4 Pos. 10).
DO NOT LEAN ON AND DO NOT REST OBJECTS ON THE FLOWMETER
BOX (BOTH WALL AND UNIT-MOUNTED VERSIONS).
DO NOT FORCE THE ARM OF THE WALL-MOUNTED VERSION.
SLOWLY OPEN THE VALVES ON THE O
VALVES COUNTER-CLOCKWISE (FIGURA 10 POS L).
Read the pressure on the pressure reducer pressure gauges, (see TECHNICAL DATA) to
estimate the autonomy of the oxygen cylinder.
Please make sure that the balloon and tubes are inserted well into their fittings.
Choose the most suitable mask for the patient.
Place the mask on the patient fitting it on well around his/her nose.
Secure the mask to the loop on the back of the head rest of the dental chair.
Make sure that the tubes for the mask circuit do not get bent or chocked.
The operator must be stationed in such a way that he/she can visually control:
-
THE RUBBER BAG (Fig. 18 Pos. F)
-
THE FLOW METERS (Fig. 18 Pos. D, E)
During the procedure constantly check the flows displayed on the flow
meters (Figure 13 Pos D, E).
Before leaving the patient, even briefly, shut off sedation.
Make sure that the %N2O knob (Fig.18 Pos C) index displays 0.
24
AND N
O CYLINDERS TURN THE
2
2