Connection To A Surgical Aspirator - Tecno-gaz Master Flux Plus Mode D'emploi

Masquer les pouces Voir aussi pour Master Flux Plus:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 69
6.4.2
UNIT-MOUNTED MASTER FLUX PLUS
Screw the hoses (Figure 2 Pos 8/9) to the unit-mounted.
Connect the mask circuit gas exhaust tube to the supplied fitting (Figura 17).
Connect the suction circuit (Figure 16) to the hose fitting.
Connect the fitting to the surgical aspirator tube (Ø11 mm).
WALL-MOUNTED MASTER FLUX PLUS
Remove the tube (Figure 16 Pos 43).
Connect the tube, gas exhaust of the mask circuit, to the valve (Figure 16 Pos 44).
Connect the fitting to the surgical aspirator tube (Ø11).
VALVE ADJUSTMENT
Turn the perforated ring nut on the valve (Figura 16 Pos 44) until the holes close.
Start gas delivery by setting ~10 l/min.
Start the surgical aspirator by setting the minimum aspiration value.
Observe the bar (Figura 13 Pos F).
Turn the perforated ring nut on the valve gradually opening the holes until the bag reaches
its ideal condition of "HALF FULL HALF EMPTY"
By changing the flow delivery it may be necessary to adjust the correct aspiration
flows of the exhaled gases.
FIGURE
POSITION
41
9A
42
43
44

CONNECTION TO A SURGICAL ASPIRATOR

ATTENTION
Figure 16 (Art MF824ZMF)
Q.TY
DESCRIPTION
1
REDUCTION
1
TERMINAL PART
2
PIPES
1
VALVE
ENGLISH
21

Publicité

Table des Matières
loading

Table des Matières