8
8.1
If cylinder pressure is very low ~10÷20 bar, proceed with changing or re-charging the
cylinders.
We recommend replacing the nitrous oxide bottle every 2 oxygen tanks.
Slowly close the valves (turn counter-clockwise).
Discharge the system by opening it from the "FLOW" and '%' knob, then close it back up.
Take the pressure reducers out, keeping them in a covered and clean location. (See
installation)
Re-charging, checking and periodic inspection of the cylinders must be carried out by
specialised companies in compliance with regulations in force in the country of use.
8.2
Use detergents that are not aggressive to plastic
Clean the surfaces of the device with a dry cloth.
Before and after use clean, wipe or sterilise the parts that come into contact with the patient
or his/her breath.
Clean the protection Plaxiglass® of the flow meters with an antistatic detergent
MASTER FLUX PLUS components cleaning table
FIG/POS
MATERIAL
3 / F
LATEX
OPTIONAL
NEOPRENE
For cleaning the mask (silicone parts and shell) refer to the "Circuits and Masks"
user manual.
THE SUPPLIED ITEMS VARY ON THE BASIS OF THE DEVICE CONFIGURATION
MAINTENANCE
CYLINDERS CHANGING
CLEANING
CODE
DESCRIPTION
3MEDN0001
BAG 3LT
3MEDN0002
BAG 3LT
ENGLISH
NUMBER OF
STEAM
SERIALIZATION
STERILISATION
CYCLES
121°
134°÷137°
15
-
29