Vercise™ DBS Leads Directions for Use

Intraoperative Testing

The External Trial Stimulator (ETS), OR Cable, OR Cable Extension, ETS Adapter, and Clinician
Programmer (CP) are used to conduct intraoperative stimulation testing during the procedure. The
OR Cable Extension is designed for temporary connection to the OR Cable to facilitate stimulation
testing outside of the sterile field. The specific model of ETS, ETS Adapter, and OR Cables to be
used depends on the Stimulator being implanted. See your DBS Reference Guide to determine
the External Trial Stimulator, ETS Adapter, and OR cables compatible with the Stimulator being
implanted.
Note: Refer to the Vercise Physician Manual for Instructions for ETS 1.
Intraoperative Testing with External Trial Stimulator 2 and 2x8 OR Cables
and Extension
The following steps are for intraoperative testing to determine lead placement using the ETS 2
(Model number DB-5132) and 2x8 OR Cable and Extension (Model number DB-4108 or SC-4108).
Refer to the appropriate Programming Manual for detailed stimulation procedures and guidelines.
CAUTION: Do not immerse the OR Cable connector or plug in water or other liquids. The
OR Cable is intended for one-time use only; do not resterilize.
1. Attach the OR Cable Extension to the OR Cable (Figure 4).
Figure 4: External Trial Stimulator 2 and 2x8 OR Cable and Extension
2. Ensure that the External Trial Stimulator 2 is off by checking the Stim Indicator light on the
device.
3. Connect the OR Cable Extension to the External Trial Stimulator 2 socket labeled "CD"
(Figure 5).
Note: The descriptors "proximal" and "distal" use the Stimulator as reference.
Vercise™ DBS Leads Directions for Use
91172963-02 REV A 4 of 329