4. Plug the OR Cable with Extension into the ETS Adapter socket labeled "1-L" (Figure 9). If two
DBS leads are being used, connect the left DBS Lead to socket "1-L" and the right DBS Lead to
socket "2-R."
Figure 9: Connection of the 1x8 OR Cable to External Trial Stimulator 2 and ETS Adapter
5. Check that the locking lever on the OR Cable connector is in the open (0) position.
6. Slide the proximal end of the DBS Lead, with the stylet, into the open port on the OR Cable
connector.
Note: The descriptors "proximal" and "distal" use the Stimulator as reference.
7. Push the end of the DBS Lead into the port until it stops. Hold the DBS Lead in place while
sliding the locking lever to the locked (1) position (Figure 10).
Figure 10: Securing the DBS Lead into the 1x8 OR Cable Connector
8. Secure the OR Cable with Extension to the microdrive or stereotactic frame.
9. Verify impedances are acceptable by using the CP or Remote Control to measure impedances.
locked
Vercise™ DBS Leads Directions for Use
91172963-02 REV A 7 of 329