Figure 5: Connection of the 2x8 OR Cable to the External Trial Stimulator 2
4. Ensure that the locking lever on the OR Cable connector is in the open (
5. Slide the proximal end of the DBS Lead, with stylet, into the open port labeled "C" on the OR
Cable connector. If two DBS leads are being used, connect the left DBS lead to Port C and the
right DBS lead to Port D.
6. Push the end of the DBS Lead into the port until it stops. Hold the DBS Lead in place while
sliding the locking lever to the locked (
WARNING: Always turn the External Trial Stimulator 2 off before connecting or
disconnecting the cable assemblies to prevent unexpected stimulation.
Figure 6: Securing the DBS Lead into the 2x8 OR Cable Connector
7. Secure the OR Cable to the microdrive or stereotactic frame.
8. Verify impedances are acceptable by using the CP or Remote Control to measure impedances.
9. Evaluate Lead placement:
(a). Evaluate whether stimulation is appropriate to reduce symptoms.
(b). Adjust the Lead location or stimulation parameters if necessary.
Note: The stylet should remain in place throughout Lead insertion and adjustments.
WARNING: High charge density can cause permanent tissue damage. The Clinician
Programmer will limit stimulation parameters to safe values.
WARNING: Increasing the number of Lead penetrations increases the probability of
hemorrhage. The necessity for an acute Lead revision should be minimized
using techniques of target localization, such as microelectrode recordings and
imaging, to correctly place the Leads on the first attempt.
) position (Figure 6).
Vercise™ DBS Leads Directions for Use
91172963-02 REV A 5 of 329
) position.