Vercise™ DBS Leads Directions for Use
10. Turn off the External Trial Stimulator 2.
WARNING: A sudden increase in stimulation may occur if External Trial Stimulator 2 is ON
while disconnecting the OR Cables.
11. Disconnect the OR Cable and OR Cable Extension from the proximal end of the Lead.
12. Verify that the Lead has not moved from the desired location.
Intraoperative Testing using the External Trial Stimulator 2, the ETS
Adapter, and 1x8 OR Cables and Extension
The following steps are for intraoperative testing to determine lead placement using the ETS 2
(Model number DB-5132) and ETS Adapter (DB-9315) and 1x8 OR Cable and Extension (Model
number DB-4100A or SC-4100A). Refer to your Programming Manual for detailed stimulation
procedures and guidelines.
CAUTION: Do not immerse the OR Cable connector or plug in water or other liquids. The
OR Cable is intended for one-time use only; do not resterilize.
1. Attach the OR Cable Extension to the OR Cable (Figure 7).
2. Verify that the External Trial Stimulator 2 is off.
WARNING: Always turn the External Trial Stimulator 2 off before connecting or
disconnecting the cable assemblies to prevent unexpected stimulation.
3. Plug the ETS Adapter into the External Trial Stimulator 2 socket labeled "CD" (Figure 8).
Figure 8: Connection of the ETS Adapter to the External Trial Stimulator 2
Vercise™ DBS Leads Directions for Use
91172963-02 REV A 6 of 329
Figure 7: 1x8 OR Cables and Extension