6. Insert the Lead, with the stylet in place, into the cannula.
7. Attach the Lead to the microdrive. You may orient the directional contacts by positioning the directional
marker (see Figure 3) in a desired position when attaching the Lead to the microdrive.
Figure 3: Directional Lead Marker
8. Slowly advance the Lead to the desired target using the microdrive.
Note: Ensure that the stylet is inside the Lead before advancing the Lead to the desired target.

Intraoperative Testing

The External Trial Stimulator, OR Cable, OR Cable Extension, and Clinician Programmer (CP) are used to
conduct intraoperative stimulation testing during the procedure. The OR Cable Extension is designed for
temporary connection to the OR Cable to facilitate stimulation testing outside of the sterile field.
The following steps are for procedural reference only. Refer to the appropriate Programming Manual for
detailed stimulation procedures and guidelines.
CAUTION: Do not immerse the OR Cable connector or plug in water or other liquids. The OR Cable is intended for
one-time use only; do not resterilize.
1. Ensure that the locking lever on the OR Cable connector is in the open (
2. Slide the proximal end of the Lead, with stylet, into the open port labeled "C" on the OR Cable
connector. If two leads are being used, connect the left lead to Port C and the right lead to Port D.
Note: The descriptors "proximal" and "distal" use the Stimulator as reference.
marker
Figure 4: OR Cable Extension
Instructions for Use
) position.
Vercise™ DBS Directional Lead