Intraoperative Testing - Boston Scientific Vercise DBS DB-2201-30AC Mode D'emploi

Masquer les pouces Voir aussi pour Vercise DBS DB-2201-30AC:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 34
Figure 2: Applying the DBS Lead Stop
6. Insert the DBS Lead, with the stylet in place, into the cannula.
7. Insert the DBS Lead and the cannula into the cannula guide on the microdrive.
8. Attach the DBS Lead to the microdrive.
Note:
Ensure that the stylet is inside the Lead before advancing the Lead to the desired target.
9. Slowly advance the DBS Lead to the desired target using the microdrive.

Intraoperative Testing

The External Trial Stimulator, OR Cable, OR Cable Extension, and Clinician Programmer (CP) are
used to conduct intraoperative stimulation testing during the procedure. The OR Cable Extension
is designed for temporary connection to the OR Cable to facilitate stimulation testing outside of
the sterile field. The External Trial Stimulator and OR Cables used depends on the Stimulator
being implanted. Intraoperative testing can be performed with either External Trial Stimulator 1 or
External Trial Stimulator 2 and compatible OR Cables, depending on the Stimulator that will be
implanted. For additional information, see your DBS Reference Guide.
Intraoperative Testing with External Trial Stimulator 1 and OR Cables
Figure 3: External Trial Stimulator 1 and OR Cables
The following steps are for procedural reference only. Refer to your Programming Manual for
detailed stimulation procedures and guidelines.
CAUTION: Do not immerse the OR Cable connector or plug in water or other liquids. The
OR Cable is intended for one-time use only; do not resterilize.
Vercise™ DBS Lead Directions for Use
90985315-02 REV B 3 of 317

Publicité

Table des Matières
loading

Table des Matières