Combined ultrasound and pneumatic lithotripsy

Assembly of the Vario Ultrasound Handpiece for
combined use with the Pneumatic Handpiece pn3
The equipment has been provided non-sterile and must be
cleaned and sterilized before use, please refer to the "Reprocessing
instructions" manual.
To thread the suction connector in properly, align it carefully
with the handpiece and apply a light axial pressure while screwing it
in place.
8
FB-389_US_ed_2009-12 User man US handpiece BSC.indd 8
The Ultrasound Probe must be tightened onto the Ultrasound
Handpiece by using the supplied wrench. Store the wrench sterile. If
the probe loosens during treatment, retighten the probe by using the
wrench. Even a slightly loose probe connection will cause a total loss
of fragmentation performance.
STERILE
6/3/2015 2:32:03 PM