Adjust the length of the Pneumatic Probe by turning the wheel of the
adjustment interface so that the tip of the Pneumatic Probe is flush
with the tip of the Ultrasound Probe.
If the Pneumatic Probe cannot be adjusted flush with the
Ultrasound Probe, this indicates the wrong ultrasound / Pneumatic
Probe combination is being used. Combination probes are marked
by an identical ring pattern on the proximal shaft of both probes.
10
FB-389_US_ed_2009-12 User man US handpiece BSC.indd 10
2
1
Both lithotripters work only in direct contact with calculi. The
length adjustment of the Pneumatic Probe relative to the Ultrasound
Probe is important, when using both lithotripters simultaneously.