TROUBLESHOOTING

Symptom
The unit does not start or stops
The foot pedal does not activate the unit
The unit is still activated when the foot
pedal is released
Hand piece inert, or unstable function
58
Solution
Check the electrical connections
Check the fuses on back panel
Check main switch function: At least one LED must be visible on adjustments and working mode areas
The unit is in "stand-by" mode, tap on any sensitive zone to exit stand-by
Send the unit to your authorized EMS Service Center
Check that the foot pedal is properly connected
Activate the pump with the back up switch on back panel
Replace the foot pedal
Send the unit to your authorized EMS Service Center
The foot pedal is active, check for jam at foot pedal
Disconnect the foot pedal and check function with the back up switch on back panel
The switch in the pedal is blocked. Turn the unit off at the power switch
Replace the foot pedal
Send the unit to your authorized EMS Service Center
Check cord connection to the unit
Verify the power setting, increase power setting to maximum
Check that the instrument is tight on the handpiece using the combitorque
Install a new instrument on the handpiece
Replace the handpiece and/or the handpiece cord
Send the unit to your authorized EMS Service Center
cont. next page