TROUBLESHOOTING
Symptom
The unit does not start or stops
The foot pedal does not activate the unit
The unit is still activated when the foot
pedal is released
Hand piece inert, or unstable function
58
Solution
•
Check the electrical connections
•
Check the fuses on back panel
•
Check main switch function: At least one LED must be visible on adjustments and working mode areas
•
The unit is in "stand-by" mode, tap on any sensitive zone to exit stand-by
•
Send the unit to your authorized EMS Service Center
•
Check that the foot pedal is properly connected
•
Activate the pump with the back up switch on back panel
•
Replace the foot pedal
•
Send the unit to your authorized EMS Service Center
•
The foot pedal is active, check for jam at foot pedal
•
Disconnect the foot pedal and check function with the back up switch on back panel
•
The switch in the pedal is blocked. Turn the unit off at the power switch
•
Replace the foot pedal
•
Send the unit to your authorized EMS Service Center
•
Check cord connection to the unit
•
Verify the power setting, increase power setting to maximum
•
Check that the instrument is tight on the handpiece using the combitorque
•
Install a new instrument on the handpiece
•
Replace the handpiece and/or the handpiece cord
•
Send the unit to your authorized EMS Service Center