Troubleshooting

Fault
The green equipment-on lamp
(3) does not illuminate when
the unit is switched on
No ultrasonic oscillations when
the footswitch (18) is activated
Possible cause
No power supply, power switch
(2) OFF
Mains plug not plugged in or
defective
Mains socket not live
The green lamp is defective
Defective fuse
Electrical fault
Power switch (2) switched
OFF
Use the first footswitch setting
with the rocker switch set to
IRRIGATION/SCALING
The handpiece hose (8) is in
the handpiece holder (9)
The instrument or file holder
have not been fixed in place
properly
The handpiece of the system
has not been fixed in place
properly
There is liquid in the
handpiece hose (8) coupling
The instrument or file holder is
damaged
The handpiece hose (8) or
handpiece is defective
Defective electronics
Piezon Master 400
Remedy
Switch the power switch (2)
ON
Check the mains plug
Check the power supply in the
practice
Send the unit to the EMS service
department
Send the unit to the EMS service
department
Send the unit to the EMS service
department
Switch power switch (2) ON
Press on the footswitch firmly
Remove the handpiece (8) from
the handpiece holder (9)
Check that the instrument or file
holder has been fixed in place
properly
Check that the handpiece has
been fixed in place properly
Dry the O-ring with compressed
air and check the handpiece
plug
Order a new instrument or file
holder
Send the unit to the EMS service
department
Send the unit to the EMS service
department