Symptom
Liquid flow too low, or instrument tends
to accumulate heat
Liquid flow too high
The pump does not work
The hand piece does not remain on the
magnetic support
The hand piece cap does not remain
tight
Solution
The pump does not function, see below
Remove the pump and check that drive shaft rotates smoothly when activated
Check that the silicon tubing is correctly placed in the pump
Check that the sterile line is the one recommended by EMS for this unit
Check that the sterile line is not damaged or obstructed
Check flow adjustment, increase flow to maximum
Check that the instrument orifice is not partially clogged up
Send the unit to your authorized EMS Service Center
Check flow adjustment and decrease it
Check that pump is correctly positioned on support
Check that the hand piece is correctly connected to the unit
Remove the pump and check if the drive shaft rotates
Check that the foot pedal is properly connected to unit
Activate the pump with the backup switch on back panel
Replace the foot pedal
Send the unit to your authorized EMS Service Center
Check the seal at the support base
Install the magnetic support on a metallic part of the unit
The support is demagnetized
Replace magnetic support
Check for presence of the O-ring seal and replace if necessary
Check the thread on handpiece
Replace the handpiece