6. MAINTENANCE & TROUBLESHOOTING
®
6.1. AIRFLOW
In case of a clogged handpiece and before the reprocessing of AIRFLOW
handpieces.
Rinse through central lumina in
the normal direction of flow (no back
rinsing) using Easy Clean with a
disposable syringe filled with more
than 2 ml of drinking water
The Easy Clean tool can be thermally disinfected and also sterilized at up to 135°C in the autoclave.
®
6.2. AIRFLOW
In case of leakage at the AIRFLOW
connection with the AIRFLOW
of the cord with the spare provided in the EL-651 Kit
located in the
®
6.3. PIEZON
light guide check & replace
The light guide loses its transparence after undergoing repeated sterilization and reprocessing
cycles. Check the transparency of the light guide every month and do the following:
ENGLISH
handpiece powder unclogging
Easy Clean FV-083
Provided in your
handpiece leakage
®
cord, replace the o-rings
application box.
®
1.
Remove the tip and unscrew the handpiece nose cap by hand.
2.
Take off the light guide and inspect it.
3.
Place in a new light guide AB-340 (provided in the
4.
Screw the nose cap on again, by hand only.
FB-618/EN – Rev. A-01 - 2017.03.30
Application box
®
Blow air to dry.
®
handpiece
®
and PERIOFLOW
®
application).
33/46
®