System Activation - Agilent Technologies TPS-mobile 969-8400 Notice De Mode D'emploi

Masquer les pouces Voir aussi pour TPS-mobile 969-8400:
Table des Matières

Publicité

ISTRUZIONI ORIGINALI
IEC320 electrical socket on the rear side of the
system (see picture). The socket is provided
with a retention spring to avoid accidental
disconnection of power cord plug – use the
spring to fasten the power cord plug.
CONNECTION TO MAIN
− Provide the correct voltage to the system
through the power cord.
− The system is equipped with a main switch
(thermal breaker) able to protect the system
components against overload or short-circuits .
Move the breaker switch to the position "1" to
start operating with the TPS-mobile.

SYSTEM ACTIVATION

(with thermal protection)
− The system activation by means of the Main
Switch provides voltage to TMP controller but
both TMP and Foreline pump continue to be
switched off.
The
system
activation
controller to start reading the pressure data
through the optional FRG-700 Full Range
Gauge if it is connected to the gauge
connection port on the rear side of the system.
Retention
Spring
Power cord
plug
NOTE
PN 969-8433 is already supplied with
an integrated power cord so you don't
need to perform this step.
Main switch
allows
the
TMP
GAUGE CONNECTION
− It is possible to start pumping by means of the
selector named "Pumping". As soon as the
Pumping selector is moved to the position "1"
the TMP and Foreline pump will be switched
on.
START PUMPING
PNs 9698404, 9698405, 9698406, 9698419,
9698430, 9698431, 9698432, 9698433, are not
equipped with the selector named "Pumping", the
same function is provided by the pushbutton
Start/Stop on the TMP controller.
TPS-mobile switching off:
− You can switch the pumps off by using the
"pumping" selector (moving it to position "0") –
the TMP controller will remain ON and if a
gauge is connected to the system it will be
possible to continue reading the pressure even
if pumps are off.
− An alternative mode for switching the pumps off
is the use of Main Switch (move it to position
"0") – The system will be completely switched
off and no reading or operation will be possible.
Emergency Stop
If an emergency situation occurs is possible to
switch the pumps and the controller off, even
disconnecting the mains cable (emergency stop).
84/105
INSTRUCTIONS FOR USE
Gauge port
Serial port
"Pumping" selector
NOTE
87-901-005-01(A)