EN-78
11.2 Appendix B
LEVEL 0
LEVEL 1
Find Quality Products Online at:
Table 5-3: Status Byte B Bit Definitions
Status Bits
Bit 0
Bit 1
Bit 2
Out of Range = 1 (Over capacity or Under Zero)
Bit 3
Bit 4
lb = 0, kg = 1 (see also Status Byte C, bits 0, 1, 2)
Bit 5
Bit 6
Table 5-4: Status Byte C Bit Definitions
Bits 2, 1, and 0
Weight Description
2
1
0
0
0
0
lb or kg, selected by Status Byte B, bit 4
0
0
1
grams (g)
0
1
0
metric tons (t)
0
1
1
ounces (oz)
1
0
0
not used
1
0
1
not used
1
1
1
tons (ton)
1
1
1
no units
Bit 3
Print Request = 1
Bit 4
Expand Data x 10 = 1, Normal = 0
Bit 5
Always = 1
Bit 6
Always = 0
Command
Function
@
Reset the scale
I0
Inquiry of all available SICS commands
I1
Inquiry of SICS level and SICS versions
I2
Inquiry of scale data
I3
Inquiry of scale software version
I4
Inquiry of serial number
S
Send stable weight value
SI
Send weight value immediately
SIR
Send weight value repeatedly
Z
Zero the scale
ZI
Zero immediately
D
Write text into display
DW
Weight display
SR
Send and repeat stable weight value
T
Tare
TA
Tare value
TAC
Clear tare
TI
Tare immediately
GlobalTestSupply
www.
Function
Gross = 0, Net = 1
Sign, Positive = 0, Negative = 1
Motion = 1, Stable = 0
Always = 1
Zero Not Captured after power-up = 1
MT-SICS Commands
Defender 5000 Indicators
.com
sales@GlobalTestSupply.com