EN-20
Press the Softkey corresponding to the icon
enter the Calibration sub-menu.
Scroll to Linearity Calibration using the Softkey
corresponding to the icon
Press the Softkey corresponding to the icon
initiate Linearity Calibration.
Clear the pan and press the Softkey corresponding to
the icon
Put the calibration mass of the specified weight on
the pan, and then press the Softkey corresponding
to the icon
different calibration point, input the value desired,
and then place the corresponding weight on the pan
for calibration.
Put the calibration mass on the pan, and then press
the Softkey corresponding to the icon
confirmation. To change to a different calibration
point, input the value desired, and then place the
corresponding weight on the pan for calibration.
The message 'Completed' will be displayed on the
screen.
Exit Linearity Calibration by pressing the Softkey
corresponding to the icon
To return to the Main Menu, press the Softkey
corresponding to the icon
3.3.4 GEO Adjustment
Set the GEO factor that corresponds to your location.
GEO codes are numbered 0-31.
Long press the button
Main Menu. Select the menu item Calibration
by pressing the Softkey corresponding to the
icon
Scroll to GEO Adjustment using the Softkey
corresponding to the icon